• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN 840 VENTILATOR; CONTINUOUS VENTILATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN 840 VENTILATOR; CONTINUOUS VENTILATOR Back to Search Results
Model Number 840
Device Problem Device Emits Odor (1425)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The service engineer (se) installed the power supply and performed est/sst/pvt and electrical safety, unit passed all test and calibrations as outlined in service procedures.
 
Event Description
It was reported that, an 840 ventilator had a burning smell.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
840 VENTILATOR
Type of Device
CONTINUOUS VENTILATOR
Manufacturer (Section D)
COVIDIEN
new mervue industrial park
michael collins road
galway
EI 
Manufacturer (Section G)
COVIDIEN
new mervue industrial park
michael collins road
galway
EI  
Manufacturer Contact
ray maroofian
2101 faraday ave
carlsbad, CA 92008
7606035334
MDR Report Key6283611
MDR Text Key66165669
Report Number8020893-2017-00338
Device Sequence Number1
Product Code CBK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K970460
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial
Report Date 01/05/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/27/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number840
Device Catalogue Number4-840120DIUU-US
Was Device Available for Evaluation? Yes
Date Manufacturer Received01/05/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/16/1998
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-