Model Number SC-1132
Device Problems Failure to Charge (1085); Telemetry Discrepancy (1629)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/14/2017
Event Type  Injury  
Event Description
A report was received that the patient was not able to charge ipg.It was noted that the ipg was having a communication difficulty.The patient will undergo a revision procedure.
 
Manufacturer Narrative
Additional information was received that the ipg was implanted too deep which caused the difficulty charging.The patient underwent a pocket revision procedure wherein the pocket was made shallow/closer to the skin.The patient was doing well postoperatively.
 
Event Description
A report was received that the patient was not able to charge ipg.It was noted that the ipg was having a communication difficulty.The patient will undergo a revision procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECISION SPECTRA®
Type of Device
SPINAL CORD STIMULATOR
Manufacturer (Section D)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer (Section G)
BOSTON SCIENTIFIC NEUROMODULATION
25155 rye canyon loop
valencia CA 91355
Manufacturer Contact
talar tahmasian
25155 rye canyon loop
valencia, CA 91355
6619494863
MDR Report Key6283927
MDR Text Key65975249
Report Number3006630150-2017-00287
Device Sequence Number1
Product Code LGW
UDI-Device Identifier08714729821526
UDI-Public(01)08714729821526(17)180829(10)19645015
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
030017
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/14/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/27/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date08/29/2018
Device Model NumberSC-1132
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/30/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/06/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age64 YR
-
-