Model Number MMT-754CAH |
Device Problems
Dent in Material (2526); Naturally Worn (2988)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/05/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.The insulin pump involved in this event is the paradigm real-time veo insulin infusion pump, which is not marketed in the united states.However, the device is similar to the paradigm real-time insulin infusion pump, which is marketed in the united states.
|
|
Event Description
|
The customer reported via phone call that the bottom of the reservoir compartment was worn out and the drive support cap was indented.The customer's blood glucose was 10.9 mmol/l at the time of incident.The customer stated that the drive support cap was flush.The customer was advised that the insulin pump will need to be replaced.The customer was advised to revert to back-up plan.The insulin pump will not be returned for analysis.
|
|
Manufacturer Narrative
|
The insulin pump was received with partially detached end cap.Drive support disk was inspected and no anomaly was noted.Device was received with cracked case at display window corner, cracked lcd window, minor scratched lcd window, cracked battery tube threads and cracked reservoir tube lip.
|
|