• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED Back to Search Results
Model Number MMT-723NAS
Device Problems Device Displays Incorrect Message (2591); Naturally Worn (2988)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/22/2017
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event Description
The customer reported via phone call that the insulin pump alarmed motor error.Customer's blood glucose was 104 mg/dl at the time of incident.Troubleshooting found that the pump also alarmed motor position encoder error.Customer indicated that the pump was worn in or around an mri machine.Customer was advised to discontinue use of the device and revert to a back-up plan per their doctor's instruction.The customer was also advised that the device would be replaced and agreed to return the product for analysis.
 
Manufacturer Narrative
The insulin pump alarmed motor error during rewind due to motor encoder signal out of phase.Unable to perform the self-test, unexpected restart alarm test, displacement test, rewind, basic occlusion test, occlusion test, prime test, and excessive no delivery test or verify operating currents and motor position encoder error alarm due to motor error alarms.The device had a cracked lcd window, cracked case at the display window corners, cracked battery tube threads and cracked reservoir tube lip.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key6284608
MDR Text Key66229487
Report Number3004209178-2017-87246
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
PMA/PMN Number
P980022
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 07/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/27/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-723NAS
Device Catalogue NumberMMT-723NAS
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/12/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received07/15/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age48 YR
Patient Weight171
-
-