MAQUET CARDIOPULMONARY AG HLM TUBING SET W/BIOLINE COATING; TUBING, PUMP, CARDIOPULMONARY BYPASS Back to Search Results
Model Number BEQ-HLS 7050 USA
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/17/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).A follow-up medwatch will be submitted when additional information becomes available.
 
Event Description
According to the customer: ¿has any center ever told you about their oxygenator being completely covered in fibrin on the arterial side of the oxygenator¿ i responded that i did not recall that description and i asked if they had experienced it and if it was a qid or hls circuit.She sent a picture of an hls circuit.(b)(4).
 
Manufacturer Narrative
(b)(4).In relation to the customer's request for information regarding fibrin deposits, the therapy application manager confirmed that increased fibrin deposits on the oxygenator membrane is a known phenomenon and is a typical observation for patients with sepsis on ecmo.Based on the investigation and trending for this issue, a systemic issue is not indicated, and the increase in fibrin deposits is a known risk associated with patients with sepsis on ecmo, therefore no further investigation or action is currently warranted in addition to continued periodic monitoring, and the complaint will be closed.
 
Event Description
Ref.:tw id # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HLM TUBING SET W/BIOLINE COATING
Type of Device
TUBING, PUMP, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
MAQUET CARDIOPULMONARY AG
rastatt
GM 
Manufacturer (Section G)
BERND RAKOW
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt
GM  
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
76437 rastatt 
4972229321
MDR Report Key6286540
MDR Text Key66358451
Report Number8010762-2017-00021
Device Sequence Number1
Product Code DWE
Combination Product (y/n)N
PMA/PMN Number
K080592
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/30/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBEQ-HLS 7050 USA
Device Catalogue Number701052794
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/17/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-