MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS, INC. EXCEL T-HANDLE; HIP INSTRUMENT/TRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS, INC. EXCEL T-HANDLE; HIP INSTRUMENT/TRIAL Back to Search Results
Catalog Number 200142000
Device Problem Crack (1135)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/10/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
The instrument is cracked.Update jan 23, 2017 - upon product evaluation, the instrument handle is cracked.
 
Manufacturer Narrative
Examination of the returned device confirms the reported observation; the handle is cracked.A complaint database search on the provided product code identified similar reports for handle damage/breakage.A previous evaluation found the t-handles are cracking due to environmental stress cracking.This is due to the combination of normal loading or usage of the instrument and exposure to chemicals they are not intended to be (non-compliant to cleaning guidelines) leading to weakening of the material over time.Based on the determination of environmental stress cracking as the root cause, no further corrective action is being pursued at this time.Continue to monitor complaints under post market surveillance (b)(4).Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXCEL T-HANDLE
Type of Device
HIP INSTRUMENT/TRIAL
Manufacturer (Section D)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC.
700 orthopaedic drive
warsaw IN 46582
Manufacturer Contact
chad gibson
700 orthopaedic drive
warsaw, IN 46581
5743725905
MDR Report Key6290016
MDR Text Key66166139
Report Number1818910-2017-11700
Device Sequence Number1
Product Code FZO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 01/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received01/31/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number200142000
Device Lot NumberSO2019557
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/18/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/02/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/31/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-