Catalog Number 200142000 |
Device Problem
Crack (1135)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 01/10/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).This complaint is still under investigation.Depuy will notify the fda of the results of this investigation once it has been completed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
|
Event Description
|
The instrument is cracked.Update jan 23, 2017 - upon product evaluation, the instrument handle is cracked.
|
|
Manufacturer Narrative
|
Examination of the returned device confirms the reported observation; the handle is cracked.A complaint database search on the provided product code identified similar reports for handle damage/breakage.A previous evaluation found the t-handles are cracking due to environmental stress cracking.This is due to the combination of normal loading or usage of the instrument and exposure to chemicals they are not intended to be (non-compliant to cleaning guidelines) leading to weakening of the material over time.Based on the determination of environmental stress cracking as the root cause, no further corrective action is being pursued at this time.Continue to monitor complaints under post market surveillance (b)(4).Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
|
|