• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL OPTICS INTRALASE FS2; FEMTOSECOND LASER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL OPTICS INTRALASE FS2; FEMTOSECOND LASER Back to Search Results
Model Number 20005D
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/05/2017
Event Type  Injury  
Manufacturer Narrative
Field service specialist visited the account and was unable to reproduce the beam steering error.He peaked power and preformed preventive maintenance and the unit passed all abbott specifications.All pertinent information available to abbott medical optics has been submitted.
 
Event Description
Account reported that system stopped during patient treatment with error message and were not able to complete the procedure that same day.Patient was re-scheduled.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTRALASE FS2
Type of Device
FEMTOSECOND LASER
Manufacturer (Section D)
ABBOTT MEDICAL OPTICS
santa ana CA
Manufacturer (Section G)
ABBOTT MEDICAL OPTICS INC.
510 cottonwood drive
milpitas CA 95035
Manufacturer Contact
valerie sedzicki
1700 east st. andrew place
santa ana, CA 92705
7142478567
MDR Report Key6291906
MDR Text Key66214875
Report Number3006695864-2017-00059
Device Sequence Number1
Product Code HNO
UDI-Device Identifier05050474573451
UDI-Public(01)05050474573451
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K060372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other
Type of Report Initial
Report Date 01/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Physician
Device Model Number20005D
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received01/05/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/31/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-