COVIDIEN PALINDR RT - PKIT 15 FR X 28CM; DIALYSIS CATHETER Back to Search Results
Model Number 8888541028P
Device Problems Leak/Splash (1354); Split (2537)
Patient Problem No Information (3190)
Event Date 11/29/2016
Event Type  malfunction  
Manufacturer Narrative
Submit date: 02/01/2017.An investigation is currently under way.Upon completion the results will be forwarded.
 
Event Description
It was reported to covidien on (b)(6) 2017 that a customer had an issue with a dialysis catheter.The customer states split catheter was placed during a fistulogram procedure and leakage was noted at the hub clamp site.Several attempts have been made to obtain additional information with no response to date.
 
Manufacturer Narrative
Submit date: 02/10/2017.An investigation was performed.This complaint has not been confirmed.The actual sample involved in the reported incident was not returned for evaluations.No additional information, pictures or videos were received.Since no sample was returned for examination, it was not possible to evaluate it as part of a comprehensive failure investigation.A device history review revealed no discrepancies that may have contributed to a complaint of this failure mode.All quality assurance testing performed during manufacturing was acceptable.The quality assurance review of the visual, physical and dimensional evaluation results indicated that the product met specification requirements.In addition, all dhr are reviewed for accuracy prior to product release.No trends or triggers have been found, therefore, a corrective or preventive action is not deemed necessary at this time.It must be noted that in-process controls (such as personnel training, incoming quality acceptance testing for raw material, 100% in process visual inspection and visual acceptance sampling are performed in the plant) are in place to prevent nonconforming product from leaving the manufacturing operations.This complaint will be used for tracking and trending purposes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PALINDR RT - PKIT 15 FR X 28CM
Type of Device
DIALYSIS CATHETER
Manufacturer (Section D)
COVIDIEN
covidien manufacturing solulfons sa
edificio 820 calle #2 zona franca coyol
alajuela
Manufacturer (Section G)
COVIDIEN
covidien manufacturing solulfons sa
edificio 820 calle #2 zona franca coyol
alajuela
CS  
Manufacturer Contact
edward almeida
15 hampshire st.
mansfield, MA 02048
5084524151
MDR Report Key6294709
MDR Text Key66509635
Report Number3009211636-2017-00100
Device Sequence Number1
Product Code MSD
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/01/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number8888541028P
Device Catalogue Number8888541028P
Device Lot Number1612500135
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Date Manufacturer Received02/10/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age62 YR
-
-