A correction was made for initial reporter¿ as (b)(6) is not a physician but rather an other healthcare professional.Additional information will be provided within this report as the device catheter of the gore® viabahn® endoprosthesis was returned to gore for evaluation.Our engineers evaluation showed following: the catheter was returned.No deployment line, knob, or endoprosthesis was returned.The portion of broken distal shaft with distal tip was not returned.The transition appears melted and bonded.There is a fragment of distal shaft material present inside the transition with approximately 20mm of insertion.The distal shaft appears broken at the tip end of the transition, which is also damaged.Based on the device examination performed, no manufacturing anomalies were identified to which the event could be definitively attributed.
|