• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY SYNTHES POWER TOOLS BLACKMAX-NEURO LOW PRESSURE; MOTOR, DRILL, PNEUMATIC - HANDPIECE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY SYNTHES POWER TOOLS BLACKMAX-NEURO LOW PRESSURE; MOTOR, DRILL, PNEUMATIC - HANDPIECE Back to Search Results
Catalog Number BLACKMAX-N-LP
Device Problems Air Leak (1008); Device Operational Issue (2914)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that during an unspecified surgical procedure, it was discovered that the motor device had an air leak and the axis was blurred.It was not reported if there were any delays in the surgical procedure or if a spare device was available.There was patient involvement reported.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.The exact date of this event was unknown.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
The actual device was returned for evaluation.Reliability engineering evaluated the device and determined that the outer hose was worn.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BLACKMAX-NEURO LOW PRESSURE
Type of Device
MOTOR, DRILL, PNEUMATIC - HANDPIECE
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key6300278
MDR Text Key66471062
Report Number1045834-2017-10319
Device Sequence Number1
Product Code HBB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K831756
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberBLACKMAX-N-LP
Device Lot NumberN/I
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/19/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received01/19/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/10/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-