• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAREFUSION, INC INFANT FLOW PLUS INFANT CPAP SYSTEM WITH SIPAP FUNCTION; VENTILATOR, NON-CONTINUOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAREFUSION, INC INFANT FLOW PLUS INFANT CPAP SYSTEM WITH SIPAP FUNCTION; VENTILATOR, NON-CONTINUOUS Back to Search Results
Model Number SIPAP
Device Problems Device Stops Intermittently (1599); Device Operates Differently Than Expected (2913)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/04/2016
Event Type  Injury  
Manufacturer Narrative
(b)(4).Results of investigation: the carefusion failure analysis laboratory received the suspect device, a sipap unit, and evaluated the device.An evaluation of the device duplicated the reported issue and isolated the issue to a low battery condition.The unit had both visual and audible alarms that occurred for at least ten minutes before the unit shut down.The unit is operating per manufacturer's specifications.
 
Event Description
The customer reported that ventilator stopped working and shut down, while in use on a patient.The screen went blank and the customer does not know whether any alarms were present.The patient was removed from the ventilator and placed on another unit.There was no patient compromise.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INFANT FLOW PLUS INFANT CPAP SYSTEM WITH SIPAP FUNCTION
Type of Device
VENTILATOR, NON-CONTINUOUS
Manufacturer (Section D)
CAREFUSION, INC
22745 savi ranch parkway
yorba linda CA 92887
Manufacturer (Section G)
CAREFUSION, INC
1100 bird center dr.
palm springs CA 92262
Manufacturer Contact
jill rittorno
22745 savi ranch parkway
yorba linda, CA 92887
MDR Report Key6301246
MDR Text Key66512072
Report Number2021710-2017-05326
Device Sequence Number1
Product Code BZD
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031745
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Health Professional
Type of Report Initial
Report Date 02/03/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/03/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSIPAP
Device Catalogue NumberSIPAP ENGL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/11/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/05/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/01/2008
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-