BOSTON SCIENTIFIC - MAPLE GROVE WATCHMAN® ACCESS SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL Back to Search Results
Model Number M635TS20060
Device Problem Leak/Splash (1354)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/13/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(6).(b)(4).Device evaluated by mfr.The complaint device was not returned for analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
 
Event Description
It was reported a hemostasis valve leak occurred.A left atrial appendage (laa) closure procedure was being performed.A watchman access system was prepared and flushed successfully.The transseptal puncture was performed and the guide wire was in position.The access system was inserted and the dilator and guide wire were removed.However, the hemostasis valve on the access system was unable to close and blood was leaking out of it.They attempted to unscrew and close the valve three times, but were unsuccessful.They exchanged the access system for a new one which was used to successfully complete the procedure.There were no patient complications and the patient's condition is stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WATCHMAN® ACCESS SYSTEM
Type of Device
SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL
Manufacturer (Section D)
BOSTON SCIENTIFIC - MAPLE GROVE
one scimed place
maple grove MN 55311
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key6309185
MDR Text Key66774216
Report Number2134265-2017-00596
Device Sequence Number1
Product Code NGV
Combination Product (y/n)N
PMA/PMN Number
P130013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 01/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/07/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2019
Device Model NumberM635TS20060
Device Catalogue NumberTS-2006
Device Lot Number0019483429
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/13/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/08/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age76 YR
Patient Weight90
-
-