Affected device returned and visually inspected.The returned device has been identified as product produced within the timeframe of our 2014 resus bag recall res 68242.First, visual inspection of the affected device demonstrated the material color of the diaphragms located at the opposite end of the manometer as green and red.We changed the color of all diaphragms to red on 26 jan 2015.Second, the affected device had a white pop-off cap below the integrated manometer.This was removed on 27 jan 2015.Third, the duckbill valve material was updated on 11 feb 2015.Review of the affected duckbill valve shows cavity designation as alphabetic "i".This cavity was used for the previous material.We conclude the affected device is from a lot number in the scope of recall res 68242.This recall was initiated on 08 may 2014 and deemed closed on 18 nov 2015.
|