• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: FISHER & PAYKEL HEALTHCARE LTD ADULT NASAL INTERFACE; CAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

FISHER & PAYKEL HEALTHCARE LTD ADULT NASAL INTERFACE; CAT Back to Search Results
Model Number OPT944
Device Problem Device Issue (2379)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/09/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Method: the complaint cannula was not returned to fisher & paykel healthcare for evaluation.Our investigation is based on information provided by the customer.We were informed that the nursing staff were not present when the incident occurred and that the patient alerted the nurse that the cannula tubing had disconnected from the part that connects to the breathing circuit.Additionally the customer confirmed that the clothing clip was not being used when the incident occurred.Conclusion: based on the customer statements, we can conclude that the cannula tubing had become disconnected from the distal swivel connector.Failure to use the clothing clip may have placed undue tension on the cannula tubing and led to the disconnection.All optiflow interfaces are inspected during production for visual defects including cracks, tears, inclusions, discolouration and stretching or deformation.Any product that fails the visual inspection is disposed of.The setup instructions in the user instructions which accompany the opt944 nasal cannula include the following steps: ensure head strap clip is attached, to prevent cannula from being pulled out of the nares.Cannula can become unattached if not used with the head strap clip.Attach tubing clip to clothing/bedding to prevent cannula from pulling off face.The user instructions also contain the following warnings/cautions: do not crush or stretch tube, to prevent loss of therapy.Failure to use the set-up described above can compromise performance and affect patient safety.
 
Event Description
A hospital in (b)(6) reported via a distributor that an opt944 nasal cannula tube became disconnected from the circuit connector during patient use.The patient called the nurse and stated that he did not know how the disconnection had occurred.There was no patient consequence.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADULT NASAL INTERFACE
Type of Device
CAT
Manufacturer (Section D)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ  2013
Manufacturer (Section G)
FISHER & PAYKEL HEALTHCARE LTD
15 maurice paykel place
east tamaki
auckland, 2013
NZ   2013
Manufacturer Contact
raymond yan
173 technology drive
suite 100
irvine, CA 92618
9194534000
MDR Report Key6313283
MDR Text Key66876995
Report Number9611451-2017-00069
Device Sequence Number1
Product Code CAT
Combination Product (y/n)N
Reporter Country CodeCA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Other
Type of Report Initial
Report Date 01/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/08/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOPT944
Device Catalogue NumberOPT944
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received01/11/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-