• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS T4 ASSAY; RADIOIMMUNOASSAY, TOTAL THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS T4 ASSAY; RADIOIMMUNOASSAY, TOTAL THYROXINE Back to Search Results
Catalog Number 12017709122
Device Problem High Test Results (2457)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/09/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
The customer initially questioned results for 1 patient tested for elecsys tsh assay (tsh), elecsys ft4 ii assay (ft4 ii), elecsys pth immunoassay (pth) and elecsys t3 (t3) that showed different results using 3 different testing methods.The customer thinks there is an interference with the roche reagents.The high ft4 ii result is what made the customer investigate other thyroid results for this patient.The tsh result was normal and the elevated ft4 ii result did not match the patient¿s diagnosis since the patient does not have hypothyroidism.Three different samples from the patient were initially sent to 3 different sites using various testing methods.Based on the data provided, the results for tsh, ft4 ii and elecsys ft3 iii (ft3 iii) were erroneous and reported outside of the laboratory.The initial erroneous results for tsh, ft4 ii and ft3 iii were previously reported on manufacturer reports 1823260-2016-01872-00, 1823260-2016-01871-00 and 1823260-2016-01873-00 respectively.The customer obtained a new sample from the patient on (b)(6) 2017 and ran thyroid tests on the cobas 6000 e 601 module at the customer site and also sent it out to an external laboratory using a siemens system.The customer thinks the results from the siemens method are correct.Based on the data provided, additional erroneous results were identified for tsh and ft4 ii.New erroneous results were identified for elecsys t4 assay (t4), t3 and pth.The erroneous results were reported outside of the laboratory.Additional information on the tsh and ft4 ii results will be sent as a supplemental to those reports.This medwatch will cover t4.Refer to medwatch with patient identifier (b)(6) for information on the t3 erroneous results and medwatch with patient identifier (b)(6) for information on the pth erroneous results.Refer to the attached data for patient results.No adverse event occurred.The cobas 6000 e 601 module serial number was (b)(4).
 
Manufacturer Narrative
The patient sample from (b)(6) 2017 was submitted for investigation.Upon further investigation of the patient sample, a streptavidin interference was confirmed.This most likely caused the high t4 results.The reagent used to perform the tests contains streptavidin.This specific interference is addressed in product labeling.For diagnostic purposes, the results should always be assessed in conjunction with the patient¿s medical history, clinical examination and other findings.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS T4 ASSAY
Type of Device
RADIOIMMUNOASSAY, TOTAL THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
Manufacturer (Section G)
ROCHE DIAGNOSTICS GMBH
sandhoferstrasse 116
na
mannheim (baden-wurttemberg) 68305
GM   68305
Manufacturer Contact
michael leslie
9115 hague road
na
indianapolis, IN 46250
3175214343
MDR Report Key6314380
MDR Text Key67211359
Report Number1823260-2017-00260
Device Sequence Number1
Product Code CDX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K961490
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/09/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number12017709122
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/17/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Age16 YR
-
-