• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - CORK ROTALINK¿ PLUS; CATHETER, CORONARY, ATHERECTOMY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - CORK ROTALINK¿ PLUS; CATHETER, CORONARY, ATHERECTOMY Back to Search Results
Model Number H749236310030
Device Problem Defective Component (2292)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/18/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluated by mfr: the device was returned for analysis.A visual and microscopic inspection of the device were performed.The brake button was pushed in and not sticking out of the advancer, when received.The advancer housing was opened to check the brake and it was found that the brake button was broken or damaged.Portion of the brake was not returned for analysis.Inspection of the remainder of the device, apart from the observed damage, revealed no other damage or irregularities.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical/procedural factors.(b)(4).
 
Event Description
It was reported that the brake defeat malfunctioned.The target lesion was located in the left anterior descending artery.A 1.50 mm rotalink¿ plus was selected for use.During the procedure, it was noticed that the brake defeat button broke.The device was manually pulled out to complete the case.No patient complications were reported and the patient's status was fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ROTALINK¿ PLUS
Type of Device
CATHETER, CORONARY, ATHERECTOMY
Manufacturer (Section D)
BOSTON SCIENTIFIC - CORK
business and technology park
model farm road
cork
EI 
Manufacturer (Section G)
BOSTON SCIENTIFIC - CORK
business and technology park
model farm road
cork
EI  
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key6314574
MDR Text Key66932397
Report Number2134265-2017-00667
Device Sequence Number1
Product Code MCX
UDI-Device Identifier08714729228363
UDI-Public(01)08714729228363(17)20181130(10)20059641
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P900056
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 01/18/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/09/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2018
Device Model NumberH749236310030
Device Catalogue Number23631-003
Device Lot Number20059641
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/24/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/18/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/14/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age73 YR
-
-