• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MED REL MEDTRONIC PUERTO RICO VIVA XT; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MED REL MEDTRONIC PUERTO RICO VIVA XT; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO Back to Search Results
Model Number DTBA1D1
Device Problem False Alarm (1013)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 12/25/2016
Event Type  malfunction  
Manufacturer Narrative
The device was not returned for analysis.However, performance data collected from the device was received and analyzed.Analysis of the device memory was performed and no anomalies were found.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the device triggered a false/positive right ventricular lead integrity alert due to numerous short ventricle to ventricle counts and two high rate non-sustained episodes.The device responded appropriately, except that true, short events tripped the right ventricular lead integrity alert.There were no signs of a true lead integrity issue.The device remains in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Corrected information: sex, date of birth.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIVA XT
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
anne schilling
8200 coral sea st ne
mounds view, MN 55112
7635052036
MDR Report Key6316933
MDR Text Key67243065
Report Number3004209178-2017-03053
Device Sequence Number1
Product Code NIK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/26/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/09/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date03/14/2015
Device Model NumberDTBA1D1
Device Catalogue NumberDTBA1D1
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received12/26/2016
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/17/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
694765 LEAD, 4592-53 LEAD, 419488 LEAD
Patient Age84 YR
-
-