The suspect medical device was not returned to the manufacturer for evaluation/investigation but to olympus medical systems corporation (omsc), (b)(4)(returned to omsc on (b)(6) 2017).The evaluation/investigation confirmed that the ceramic insulation at the distal end of the inner sheath broke off completely.Furthermore, there is a lateral crack through the ceramic insulation and its inner surface shows signs of laser-induced abrasion.The cause of this damage and the breakage of the ceramic insulation is thermal overload (excessive high-frequency output power or laser-induced).It is clearly stated as a warning note in the instructions that impact, fall, shock or similar stress can damage the ceramic insulation at the sheath's distal end and that the instrument must not be used if damaged since otherwise this can cause injuries to the patient and/or user.The user apparently did not follow these instructions since the damage and breakage of the ceramic insulation were caused by thermal overload.Therefore, this event/incident was attributed to abnormal use/off-label use and the case will be closed from olympus side with no further actions.However, the event/incident will be recorded for trending and surveillance purposes.In addition, the user will be retrained to correctly use the olympus medical devices.
|