CAREFUSION ALARIS EXTENSION SET; SET, EXTENSION, INTRAVASCULAR Back to Search Results
Model Number 20028E
Device Problems Burst Container or Vessel (1074); Crack (1135); Fluid/Blood Leak (1250)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
The affected product has been received and the investigation is pending.A follow up report will be submitted once the failure investigation has been completed.
 
Event Description
The customer reported that the smartsite hub cracked when applied to primary line and fluids leaked.There is no report of patient harm.
 
Manufacturer Narrative
(b)(4).The customer¿s report that the extension set hub cracked and leaked was confirmed.During visual inspection it was observed that the female luer had a vertical hair line crack that measured 0.615 inches long.The female luer was inspected under magnification and no stress marks were observed.Functional testing confirmed leaking through the crack.The root cause of the crack was not identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALARIS EXTENSION SET
Type of Device
SET, EXTENSION, INTRAVASCULAR
Manufacturer (Section D)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer (Section G)
CAREFUSION
10020 pacific mesa blvd
san diego CA 92121 4386
Manufacturer Contact
ade ajibade
10020 pacific mesa blvd
san diego, CA 92121-4386
8586172000
MDR Report Key6330622
MDR Text Key67439061
Report Number9616066-2017-00272
Device Sequence Number1
Product Code FPB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K801614
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,other
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 02/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/14/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number20028E
Device Catalogue Number20028E
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/13/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received04/03/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-