Catalog Number 900-600
Device Problems Bent (1059); Device Issue (2379)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/30/2017
Event Type  malfunction  
Manufacturer Narrative
The reported defective device has yet to be returned to the manufacturer.An investigation into the root cause for event is currently in progress.The results of the device evaluation will be sent via a follow up medwatch.(b)(4).
 
Event Description
As reported (b)(6) 2017: during the procedure the applicator was removed for reinsertion.At the time of the second insertion the ceramic tip bent inside the patient's kidney.The tip did not detach from the applicator shaft and remained fully attached.The device was set aside and the procedure was completed with a new of the same device.It was reported the disposable device is available for return to the manufacturer for evaluation.
 
Manufacturer Narrative
Returned for evaluation was one pmta accu2i standard applicator.A visual examination of the device noted the tubing was cut and the tip was bent but still fully attached.The site of the bend occurs where the tip meets the shaft of the device.The thermocouple wires are still intact within the tip.Functional testing could not be performed due to the condition of the returned device.The customer's reported complaint description of the tip of the applicator being bent is confirmed.Although the complaint description is confirmed, a definitive root cause for the event cannot be determined.A possible contributing factor could have been operational context; i.E.Probe placement into hard tissue or lateral forces on the applicator tip during placement or removal.The instructions for use, which is supplied to with catalog number, contains a statement "always advance the applicator into the target tissue using axial forces only.Avoid placing lateral forces on the applicator tip during placement or removal" and lateral forces on the applicator tip should be avoided both during insertion and removal.Failure to do so could result in damage to the microwave array, failure of the applicator and injury to the patient." a review of the device history records was performed for the reported packaging and component lots for any deviations related to the reported defect of the complaint.The review confirms that the lots met all material, assembly, and performance specifications.A review of the angiodynamics complaint system noted no trends for this complaint type and product family.This type of complaint will continue to be monitored for trends.Complaint # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PMTA ACCU2I STANDARD APPLICATOR
Type of Device
MICROWAVE APPLICATOR
Manufacturer (Section D)
ANGIODYNAMICS
603 queensbury avenue
queensbury NY 12804
Manufacturer (Section G)
ANGIODYNAMICS, INC.
603 queensbury avenue
queensbury NY 12804
Manufacturer Contact
law ryan
10 glens falls technical park
glens falls, NY 12801
5187424488
MDR Report Key6332332
MDR Text Key67494064
Report Number1319211-2017-00024
Device Sequence Number1
Product Code NEY
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K122762
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,foreign,health pr
Reporter Occupation Medical Equipment Company Technician/Representative
Type of Report Initial,Followup
Report Date 08/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/15/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date03/31/2018
Device Catalogue Number900-600
Device Lot Number16120512
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/24/2017
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received01/30/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/05/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-