• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC - COSTA RICA (COYOL) OPTICROSS¿ 18; CATHETER, ULTRASOUND, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC - COSTA RICA (COYOL) OPTICROSS¿ 18; CATHETER, ULTRASOUND, INTRAVASCULAR Back to Search Results
Model Number H7493932800180
Device Problem Entrapment of Device (1212)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 02/01/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Device evaluated by mfr.: the device was not returned for analysis.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
 
Event Description
It was reported that guidewire entanglement occurred.During percutaneous transluminal angioplasty (pta), an opticross¿ 18 imaging catheter was advanced for visualization of the target lesion that was located in the anterior tibial artery.The guidewire was wrapped around the ivus, which was apparently spinning.The device was completely removed from the patient's body by pulling everything out.The procedure was completed with another of the same device and a new wire.No patient complications were reported and the patient's status was fine.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTICROSS¿ 18
Type of Device
CATHETER, ULTRASOUND, INTRAVASCULAR
Manufacturer (Section D)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS 
Manufacturer (Section G)
BOSTON SCIENTIFIC - COSTA RICA (COYOL)
2546 first street
propark free zone
alajuela
CS  
Manufacturer Contact
sonali arangil
one scimed place
maple grove, MN 55311
7634941700
MDR Report Key6336000
MDR Text Key67599690
Report Number2134265-2017-01287
Device Sequence Number1
Product Code ITX
Combination Product (y/n)N
PMA/PMN Number
TBD
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 02/01/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/16/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/27/2017
Device Model NumberH7493932800180
Device Lot Number19770786
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/01/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/27/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-