• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER MANUFACTURING B.V. BIPOLAR METAL SHELL 46 MM O.D.; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER MANUFACTURING B.V. BIPOLAR METAL SHELL 46 MM O.D.; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problems Bent (1059); Detachment Of Device Component (1104); Difficult to Insert (1316); Device Damaged Prior to Use (2284)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/19/2017
Event Type  malfunction  
Manufacturer Narrative
Current information is insufficient to permit a conclusion as to the cause of the event.
 
Event Description
During a procedure, there was difficulty assembling the acetabular cup with the liner on the back table.After attempting to assemble two different liners, it was discovered that the locking ring was bent inside of the acetabular cup and tabs in the cup were not aligned properly.Another cup and liner were used to complete the procedure without a significant delay.
 
Manufacturer Narrative
(b)(4).Concomitant medical products: 00500104428, bipolar liner, 63181031, 00500104428, bipolar liner, 63431201, 00801802802, versys femoral head, 63482107, 00801802802, versys femoral head, 63482109.Complaint sample was evaluated and the reported event was confirmed.A bipolar shell, and two bipolar liners were returned for evaluation.As returned, one of two tabs of the lock ring is seized in the lock ring groove of the bipolar shell.Both liners exhibit damage that indicates the lock ring tab was not in its¿ intended position prior to the attempt to install the liners.Dimensional readings of the lock ring and lock ring groove are conforming to print specifications.Dhr was reviewed and no discrepancies relevant to the reported event were found.Root cause was unable to be determined.A summary of the investigation was sent to the complainant conveying proper surgical technique and use of the device.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPOLAR METAL SHELL 46 MM O.D.
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer (Section G)
ZIMMER MANUFACTURING B.V.
turpeaux industrial park
route #1 km 123.4 bldg #1
mercedita PR 00715
Manufacturer Contact
christina arnt
56 e. bell dr.
warsaw, IN 46582
5745273773
MDR Report Key6337741
MDR Text Key67655509
Report Number0002648920-2017-00062
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
PK833991
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/09/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/16/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Physician
Device Expiration Date07/31/2023
Device Model NumberN/A
Device Catalogue Number00500104600
Device Lot Number62430399
Other Device ID Number(01) 00889024115347
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/14/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/08/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/03/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age91 YR
-
-