It was noticed during cleaning that a transducer after a prostate biopsy that blood had collected where the handle meets the transducer, it has what looks like a ring or ridge all the way around the device.After cleaning the transducer we inspected it under a magnifying lens and could see pieces of what looks like a seal is missing all round this ring.We took the device out of service, contacted our radiology department, infection control and the specialty clinic nurse manager.Ge was contacted.Pictures were taken with a microscope and sent to ge, which ge acknowledged this was an issue and replaced the devices.Two new transducers from ge were received and upon inspection these devices had the same issue.Ge was again notified and pictures sent.Ge confirmed that another two new transducers will be sent.These devices also had the same issue.It is important to note that the concern with the seal area was not visible to the naked eye and therefore would go undetected unless there was close inspection using a microscope.It was also noted that the transducer had a sharp area where the needle guide attached to the probe.As a result of this initial facility report, (b)(6) action was taken to remove these transducers from service to prevent the potential for patient harm.The investigation is ongoing, but initial findings have identified approximately (b)(4) devices with seal separation and (b)(4) had debris identified in the gap.The seal separation provides a space for bioburden to remain, possibly even after reprocessing, which poses a serious public health risk.
|