BOSTON SCIENTIFIC - COSTA RICA (HEREDIA) ROTAWIRE¿ AND WIRECLIP¿ TORQUER; CATHETER, CORONARY, ATHERECTOMY
|
Back to Search Results |
|
Model Number H802228240022 |
Device Problem
Kinked (1339)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 12/15/2016 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Device evaluated by mfr.: investigation completed.A visual inspection revealed that the device has wire broken in the middle of the device (the distal tip was returned) 323cm from the proximal section, also the wire is kinked in the middle of the device and near to the section broken, also the distal tip was found kinked.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.The most probable root cause is operational context as device performance was limited due to anatomical procedural factors.(b)(4).
|
|
Event Description
|
Reportable based on device analysis completed on (b)(6) 2017.It was reported that a rotawire kink occurred.The target lesion was located in the severely calcified coronary artery.A 330cm rotawire was selected for use.During introduction, it was noted that resistance was felt during insertion of rotalink.When the wire was checked, it was found to be kinked.The procedure was completed with another of the same device.No patient complications were reported.However, device analysis revealed that the rotawire was broken in the middle of the device.
|
|
Search Alerts/Recalls
|
|
|