Model Number FUSION COMPACT |
Device Problems
Computer Software Problem (1112); Device Stops Intermittently (1599)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 01/22/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
On 02/18/2017, a medtronic representative performed a navigation system check-out, all areas passed.System performed as intended.No parts have been received by manufacturer for analysis.No further issues have been reported.
|
|
Event Description
|
A site registered nurse (rn) reported that, while in an ear, nose & throat (ent) procedure, their navigation system was acting slow on multiple pages: in patient import page and in registration page.Noted was difficulty in verifying the probe.No further details regarding this issue were provided.The surgeon opted to complete the procedure without the use of the navigation system.Delay in therapy was greater than one hour before continuing.There was no impact on patient outcome.
|
|
Manufacturer Narrative
|
Medtronic personnel reported that the instrument the site was attempting to verify was outside of the emitter field.The site has been instructed on proper emitter placement.The software investigation found that the reported event was unrelated to a software issue.The software functioned as designed.
|
|