A hologic contura balloon applicator for hdr breast brachytherapy was found to have questionable channel problems after it was inserted into a pt.The 5 channel applicator was discovered to have the inherent radiopaque marker on channel 4 rather than being on channel 1.This resulted in an inconsistency between the identification of the channels on the brachytherapy treatment plan and the mfr's labeling on the applicator.The plan was delivered to the pt for one of 10 planned fractions before this was identified.The applicator was then removed from the pt by the attending physician.Only one treatment was delivered.Visual inspection of the applicator confirmed the marker was on channel 4.The result is that the delivered radiation dose to the pt was rotated from planned dose distribution by 90 degrees.The plan was recalculated to match how it was treated.The actual dose to the pt differed from the planned dose by less than 20% for the target volume and all organs.
|