The patient was undergoing a thrombectomy procedure in the m1 section of the internal carotid artery (ica) using a penumbra system ace 68 hi-flow kit (kit).It was noted that the patient's anatomy was tortuous.During the procedure, the physician placed a non-penumbra 9f short sheath and a non-penumbra 9f balloon catheter and then advanced the penumbra system ace 68 reperfusion catheter (ace 68) into the patient.While passing around the tortuous arch, the balloon catheter kinked and caused the ace 68 to become caught.Consequently, the ace 68 stretched and unraveled when the physician attempted to advance and retract it.Therefore, the ace 68 was removed.The physician then used a non-penumbra stent device to remove thrombus and completed the procedure.The patient was alive after the procedure; however, on the following day, the patient had another stroke in a different hemisphere and passed away.The physician does not attribute the patient's death to the penumbra device.
|
Results: the penumbra system ace 68 reperfusion catheter (ace 68) was fractured approximately 126.0 cm from the hub and stretched from approximately 104.0 cm from the hub up to the fracture.Conclusions: evaluation of the returned device revealed that the ace 68 was stretched and fractured.These types of damages are likely a result of forceful manipulation against resistance.It is likely that when the balloon catheter kinked, it pinned the ace 68 inside the balloon catheter lumen and was likely the source of resistance felt by the physician.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|