It was reported that during the treatment of a pseudoaneurysm of the right internal iliac artery, the embolization coil was positioned in place.Upon detachment, the coil did not detach.The device was pulled to withdrawal, however the delivery pusher stretched and broke.A portion of the stretched delivery pusher segment was removed, but the other portion remained in the vessel.The patient was reported to be okay.Additional information/clarification received on 4 feb 2017, indicated the following: the embolization coil was within the aneurysm.Attempts were made to remove the device by pulling.It was stated that the physician pulled the device very hard.These attempts were unsuccessful.The stretched portion of the device (delivery pusher) was left in place, no other intervention was taken.It was reported that the physician emphasized that he tried to remove the device with too much force.It was indicated that it was not the fault of the device.The patient was reported to be okay with no negative consequences.
|