SYNTHES OBERDORF BATTERY REAMER/DRILL II HANDPIECE; MOTOR, SURGICAL INSTRUMENT, AC-POWERED Back to Search Results
Catalog Number 530.705
Device Problems Mechanical Problem (1384); Loss of Power (1475)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/01/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that during an unspecified surgical procedure, it was observed that the battery reamer/drill hand piece device was powerless.It is not reported if there was a delay in the procedure due to the event, or if a spare device was available for use.There was patient involvement reported.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
Additional narrative: the device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and observed that the control unit was not functioning and was defective.It was noted that the motor power was too low and the controller sets off.It was also noted that the device failed pre-repair diagnostic tests for power at the motor with test bench.Therefore, the reported condition was confirmed.The assignable root cause was determined to be due to wear from normal use and servicing over time.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY REAMER/DRILL II HANDPIECE
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH443 6
SZ  CH4436
Manufacturer (Section G)
SYNTHES OBERDORF
eimattstrasse 3
oberdorf CH443 6
SZ   CH4436
Manufacturer Contact
michael cote
1302 wrights lane east
west chester, PA 19380
6107195000
MDR Report Key6361276
MDR Text Key68416620
Report Number8030965-2017-10965
Device Sequence Number1
Product Code GEY
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number530.705
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/24/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received02/24/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/20/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-