Catalog Number CRANI-A
Device Problem Component Falling (1105)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/01/2017
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).Reporter¿s phone number: (b)(6).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported from (b)(6) that during an unspecified surgical procedure, it was discovered that the bearing was failing on the craniotome device.It was reported that the event occurred during use on a patient.It was not reported if there were any delays in the surgical procedure or if a spare device was available.There was patient involvement reported.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Manufacturer Narrative
This device was returned for service; however, did not meet manufacturing specifications during pre-repair assessment.Reliability engineering evaluated the device and the reported condition was confirmed.It was determined that the bearings and the bracket were damaged.It was further determined that the ball bearing was worn.It was further determined that the device failed for visual assessment vibration and for temperature.The assignable root cause was determined to be due to wear from normal use and servicing.If additional information should become available, a supplemental medwatch report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
6.5CM ADULT CRANI ATTACHMT
Type of Device
MOTOR, DRILL, ELECTRIC - CRANIOTOME
Manufacturer (Section D)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer (Section G)
DEPUY SYNTHES POWER TOOLS
4500 riverside drive
palm beach gardens FL 33410
Manufacturer Contact
rodney crain
4500 riverside drive
palm beach gardens, FL 33410
5616271080
MDR Report Key6361380
MDR Text Key68420173
Report Number1045834-2017-10511
Device Sequence Number1
Product Code HBC
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K011444
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/13/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberCRANI-A
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/06/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/06/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/10/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-