Model Number CI-1500-04
Device Problem Cut In Material (2454)
Patient Problem No Code Available (3191)
Event Date 02/07/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
The recipient reportedly underwent repositioning surgery.During surgery, the electrode array was accidentally cut.The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.
 
Manufacturer Narrative
Correction: other relevant history.The recipient reportedly underwent revision surgery to improve device placement.
 
Manufacturer Narrative
(b)(4).The external visual inspection revealed the electrode array was severed prior to receipt.This is believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented one of the electrical tests from being performed.The device passed all of the electrical and mechanical tests performed.This device was explanted for medical reasons.The device passed the tests performed.This is the final report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES 90K¿ ADVANTAGE IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
Manufacturer Contact
rachael perez
28515 westinghouse place
valencia, CA 91355
6613627734
MDR Report Key6366927
MDR Text Key68595949
Report Number3006556115-2017-00085
Device Sequence Number1
Product Code MCM
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Reporter Occupation Other
Type of Report Initial,Followup,Followup,Followup
Report Date 02/07/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date12/31/2017
Device Model NumberCI-1500-04
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/29/2017
Is the Reporter a Health Professional? No
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/25/2015
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age49 YR
-
-