Model Number 2088TC/46
Device Problems Positioning Failure (1158); Kinked (1339)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 01/23/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that during the implant procedure when the lead was placed in right atrium there was kink to the ring electrode.The helix mechanism would not deploy to fix the lead.The lead was replaced without complications.The patient remained stable before, during and after the procedure.
 
Manufacturer Narrative
The damage found was sustained during the surgical procedure.The lead was otherwise normal.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TENDRIL STS
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6367997
MDR Text Key68758126
Report Number2938836-2017-16630
Device Sequence Number1
Product Code NVN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P960013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/23/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/01/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/31/2019
Device Model Number2088TC/46
Device Lot NumberA000032490
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/15/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/20/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/28/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age54 YR
-
-