• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) ELLIPSE DR ICD, US; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE) ELLIPSE DR ICD, US; IMPLANTABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number CD2377-36QC
Device Problem Difficult to Interrogate (1331)
Patient Problem Ventricular Tachycardia (2132)
Event Date 01/11/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the patient experienced ventricular tachycardia and presented at the hospital for a vt ablation procedure.The device was unable to establish communication with the programmer.The device was explanted and replaced.The patient was stable during and after the procedure.
 
Manufacturer Narrative
The reported event of non-communication was confirmed and due to low battery voltage.Final analysis found the device was prematurely depleted.High current was observed during bench testing and was traced to the hybrid.The cause of the premature battery depletion was due to excess current on the hybrid.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELLIPSE DR ICD, US
Type of Device
IMPLANTABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SUNNYVALE)
645 almanor avenue
sunnyvale CA 94085
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6374558
MDR Text Key68963074
Report Number2938836-2017-16974
Device Sequence Number1
Product Code LWS
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/29/2018
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/02/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/30/2016
Device Model NumberCD2377-36QC
Device Lot Number4789096
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/02/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/26/2018
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/01/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age61 YR
-
-