Model Number 383401
Device Problem Self-Activation or Keying (1557)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/12/2016
Event Type  malfunction  
Manufacturer Narrative
Submit date: 03/2/2017.As part of our manufacturing process, all device history records are reviewed and approved by quality, prior to release of product.The unit was sent to a service center for repair.Upon triage, the technician was unable to duplicate the customer's complaint.No fault found.The unit was cleaned, updated, tested and recertified per procedure.
 
Event Description
It was reported to covidien on (b)(6) 2016 that an issue occurred with an enteral feeding pump.The customer states the unit turns on by itself.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO JOEY,PUMP W/CLMP,REFB
Type of Device
ENTERAL FEEDING PUMP
Manufacturer (Section D)
COVIDIEN
15 hampshire st
mansfield MA 02048
Manufacturer (Section G)
COVIDIEN
mansfield
mansfield MA 02048
Manufacturer Contact
edward almeida
15 hampshire st
mansfield, MA 02048
5084524151
MDR Report Key6374917
MDR Text Key69211922
Report Number1282497-2017-00245
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 10/17/2016
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/02/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number383401
Device Catalogue Number383401
Was Device Available for Evaluation? Yes
Date Manufacturer Received10/17/2016
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-