ST. JUDE MEDICAL, INC.(CRM-SYLMAR) ZEPHYR XL DR; IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number 5826
Device Problems Device Displays Incorrect Message (2591); Programming Issue (3014)
Patient Problem Muscle Stimulation (1412)
Event Date 01/28/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
It was reported that the patient experiencing pocket stimulation presented in clinic.A merlin programmer was not present at the clinic to interrogate the device.Once a programmer was available and the device was interrogated, it exhibited an error message regarding programing parameter change error.It was suspected that a device software download was performed incorrectly causing the device to exhibit evvi.The device was reprogrammed successfully and the patient no longer experienced pocket stimulation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZEPHYR XL DR
Type of Device
IMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
denise johnson
5050 nathan lane north
plymouth, MN 55442
6517564470
MDR Report Key6381335
MDR Text Key69384389
Report Number2017865-2017-01455
Device Sequence Number1
Product Code LWP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P880086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Physician
Type of Report Initial
Report Date 01/28/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/06/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/31/2015
Device Model Number5826
Device Lot Number4342562
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received01/28/2017
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured11/12/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age82 YR
-
-