A discordant, falsely elevated creatinine_2 result was obtained on a patient sample on an advia 2400 instrument.It is unknown if the initial result was reported out to the physician(s).The customer repeated the same sample on the same advia 2400 instrument, resulting lower.It is unknown which repeat result was reported out to the physician(s).There are no known reports of patient intervention or adverse health consequences due to the discordant, falsely elevated creatinine_2 results.
|
The initial mdr 2432235-2017-00175 was filed on march 7, 2017.Additional information (05/08/2017): additional information showed that another mdr was filed for the advia 2400, serial number (b)(4).This mdr (2432235-2017-00175) is a duplicate mdr entry.For additional information, please see mdr 2432235-2017-00171.
|