As reported; patient with calcified bicuspid valve.Physicians could not cross the annulus with the device.A few attempts were performed and decision was made to change the wire.During the removal of the device, the wire got stuck in the system.When force was applied, the nose cone detached from the device.The implanters made the decision not to continue the procedure and consider another way of treatment.
|
This is a follow-up to the previous medwatch report.The manufacturing batch lot was provided.A review the device history records for the guidewire involved in this incident revealed the product met specification prior to shipment.There was no indication of manufacturing defect or anomaly that could have impacted the event as reported.The device was returned for analysis.As received, the specimen consisted of one each safari2 275 cm x sml crv; returned coiled, loose and double-bagged within "zip-lock" style poly biohazard pouches.The complaint of "the wire got stuck in the system" and "when force was applied the nosecone detached" was confirmed.The returned specimen presented ptfe coating damage scattered over the entire length of the specimen, a distorted preformed region with extensive offset/overlapping coil wrap damage.The specimen also presented multiple bends and offset/overlapping, stretched and cut coil damage located 52.85 to 127.65 cm from the proximal end with exposed core wire located 92.7 to 94.75 cm from the proximal end.The damaged lotus nosecone remained lodged on the wire shaft 94.75 to 97.15 cm from the proximal end.It was it is not possible to assign a definitive root cause for the event as reported.Based on the evidence presented by the sample and the information provided by the supporting documentation, device interface and clinical factors appear to have impacted on the event as reported.
|