• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - ENGLEWOOD EVA DUAL CHAMBER CONTAINER 3L; CONTAINER, I.V.

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - ENGLEWOOD EVA DUAL CHAMBER CONTAINER 3L; CONTAINER, I.V. Back to Search Results
Catalog Number H938905
Device Problems Fluid/Blood Leak (1250); Cut In Material (2454)
Patient Problem No Patient Involvement (2645)
Event Date 02/07/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that an exactamix dual chamber bag had a tear in the bar separating the two chambers.The small chamber then began to leak when mixing the two.This occurred during set up.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
The device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional leak testing was performed and revealed a leak on the back of the bag in the upper left corner of the lower chamber.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was verified.The cause of the condition could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVA DUAL CHAMBER CONTAINER 3L
Type of Device
CONTAINER, I.V.
Manufacturer (Section D)
BAXTER HEALTHCARE - ENGLEWOOD
englewood CO
Manufacturer (Section G)
BAXTER HEALTHCARE - ENGLEWOOD
14445 grasslands dr
englewood CO 80112
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key6387377
MDR Text Key69362264
Report Number1416980-2017-01971
Device Sequence Number1
Product Code KPE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K960581
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 04/10/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/08/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date07/06/2021
Device Catalogue NumberH938905
Device Lot Number63630-A1761
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/28/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/05/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-