Catalog Number H938905 |
Device Problems
Fluid/Blood Leak (1250); Cut In Material (2454)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 02/07/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that an exactamix dual chamber bag had a tear in the bar separating the two chambers.The small chamber then began to leak when mixing the two.This occurred during set up.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
The device was received for evaluation.Visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional leak testing was performed and revealed a leak on the back of the bag in the upper left corner of the lower chamber.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.The reported condition was verified.The cause of the condition could not be determined.Should additional relevant information become available, a supplemental report will be submitted.
|
|