Investigation has shown that the reported system behavior occurred during patient positioning and not during treatment.The sporadic, vertical txt table movement was due to a defective relay with abnormal wear on the contacts.The relay was replaced and the system in functional.Per design an uncommanded vertical table motion is detected by the system after 2mm of table motion and triggers a motion stop interlock that immediately activates the magnetic brakes to stop the movement after 2 to 3cm, which is what occurred in this case.Therefore, no risk to the patient could be identified and the system behavior is considered to be not safety related.Considering this, no further corrective action is initiated.
|