• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP; INSULIN INFUSION PUMP / SENSOR AUGMENTED Back to Search Results
Model Number MMT-751NAL
Device Problems Device Stops Intermittently (1599); Naturally Worn (2988)
Patient Problem Hyperglycemia (1905)
Event Date 03/02/2017
Event Type  Injury  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that they were hospitalized due to high blood glucose of 569 mg/dl.The customer stated that the down button is worn down and not consistent.The customer will call back to troubleshoot.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
The pump was received with all operating currents within specification and passed the functional testing including the rewind, basic occlusion, occlusion, prime, excessive no delivery, self-test, unexpected restart and displacement tests.No excessive no delivery alarms were noted.The pump passed the delivery accuracy test.The pump was received with intermittent buttons due to moisture damage at the keypad traces.The pump was received with a cracked case at the display window corners and display window.The pump was received with a stained end cap sticker, a stripped battery cap coin slot, minor scratches on the display window and case.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP
Type of Device
INSULIN INFUSION PUMP / SENSOR AUGMENTED
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50
road 31 km 24.4
juncos PR 00777 3869
Manufacturer Contact
gerwin de graaff
18000 devonshire street
northridge, CA 91325-1219
8185764805
MDR Report Key6395910
MDR Text Key69614987
Report Number3004209178-2017-94910
Device Sequence Number1
Product Code OYC
Combination Product (y/n)N
PMA/PMN Number
120010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Patient
Type of Report Initial,Followup
Report Date 03/20/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/10/2017
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberMMT-751NAL
Device Catalogue NumberMMT-751NAL
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/14/2017
Is the Reporter a Health Professional? No
Date Manufacturer Received03/20/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Other;
Patient Age26 YR
Patient Weight90
-
-