The pump was received with all operating currents within specification and passed the functional testing including the rewind, basic occlusion, occlusion, prime, excessive no delivery, self-test, unexpected restart and displacement tests.No excessive no delivery alarms were noted.The pump passed the delivery accuracy test.The pump was received with intermittent buttons due to moisture damage at the keypad traces.The pump was received with a cracked case at the display window corners and display window.The pump was received with a stained end cap sticker, a stripped battery cap coin slot, minor scratches on the display window and case.
|