• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN KANGAROO; PUMP, INFUSION, ENTERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN KANGAROO; PUMP, INFUSION, ENTERAL Back to Search Results
Model Number 382400
Device Problem Charred (1086)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Submit date: 3/10/2017.
 
Event Description
The customer states damaged rear case.Upon triage the technician has found a burnt component.
 
Manufacturer Narrative
An evaluation of the kangaroo pump was performed for the reported condition that pump had burnt component.The unit was triaged and the reported issue was confirmed.Liquid ingress caused the pcba to corrode, and is the result of customer misuse.All device history records are reviewed for quality inspections and parameter compliance prior to releasing the product for shipment.Complaint trending information is being reviewed on a monthly basis and if a trend is observed, actions will be taken as necessary.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KANGAROO
Type of Device
PUMP, INFUSION, ENTERAL
Manufacturer (Section D)
COVIDIEN
building 10- no 789 puxing roa
shanghai 20111 4
CN  201114
Manufacturer (Section G)
COVIDIEN
building 10- no 789 puxing roa
shanghai 20111 4
CN   201114
Manufacturer Contact
edward almeida
15 hampshire street
mansfield, MA 02048
5084524151
MDR Report Key6396326
MDR Text Key69745949
Report Number3006451981-2017-05054
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 09/11/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/10/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number382400
Device Catalogue Number382400
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received02/24/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-