Results: the benchmark 6f 071 delivery catheter (benchmark dc) was kinked approximately 6.0 cm from the hub.The distal tip of the benchmark dc was ovalized approximately 99.5 cm from the distal tip.Conclusions: evaluation of the returned device revealed that the benchmark dc was kinked and ovalized.These types of damages likely occurred due to forceful handling of the device during removal from the packaging.Penumbra catheters are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
During preparation for a medical procedure, the radiologist noticed that the tip of the benchmark 6f 071 delivery catheter (benchmark dc) was bent upon removal from the packaging.It was reported that the outer packaging of the device did not appear to be damaged.The bent benchmark dc was noticed prior to use and therefore, was not used in the procedure.The procedure was completed using a new benchmark dc.
|