It was reported that the patient presented to the hospital, and the device was found to be in backup operation and could not restored.Due to bradycardia, temporary pacemaker was used.There were no adverse consequences to the patient.It was later reported that upon the revision, the device could not be interrogated with the programmer and premature battery depletion was suspected.The device replacement was performed as scheduled and there were no adverse consequences to the patient during and post procedure.
|
Premature battery depletion was confirmed by analysis.Bench testing on the device was performed, and no sources of high current were noted.In further analysis of the battery, lithium clusters were observed, but at the time of analysis the clusters did not appear to be in a location or size that would be sufficient to cause an internal short of the battery.As a result, the cause of the premature battery depletion could not be conclusively determined.However, from these analyses, in the absence of other root causes, it is probable that the premature battery depletion was caused by a lithium cluster induced short circuit.The device is included in the premature battery depletion with implantable cardioverter defibrillator advisory notice issued by st.Jude medical on (b)(6) 2016.
|