Report source correction: unticked 'foreign' as report source.(b)(4).Method: one complaint 900mr810 adult heated wall reusable breathing circuit was received at fph in (b)(4) for evaluation.It was visually inspected and electrical resistance tested.Results: visual inspection revealed that the tubing film was stretched and damaged approximately midway between the breathing circuit cuffs.No melting of the tube or heaterwire was observed.The electrical resistance of the heaterwire was within specification.Conclusion: based on our investigation, the observed damage most likely occurred from the tube being stretched.All 900mr810 adult heated wall reusable breathing circuits are visually inspected and pressure tested prior to being released for distribution.Any circuits that fail are rejected.This suggests the reported damage occurred after the subject breathing circuit was released for distribution.The user instructions that accompany the 900mr810 adult heated wall reusable breathing circuit state the following: - "inspect circuit before re-use, do not use if the circuit shows signs of deterioration, such as: cracks, tears, or damage." - "perform a pressure and leak test on the breathing system, and check for occlusions, before connecting to a patient." - "do not cover the circuit with materials such as blankets, towels or bed linen." - "disconnect tube by handling end connectors only, do not pull or twist tubing, as this may cause damage.".
|