Evaluation summary: the stent was positioned on the balloon between the marker bands as per specifications.There was deformation to the 16th distal stent wrap with struts raised.Crystallized residue was present inside the inflation lumen.The device was placed in the waterbath to disperse the residue.On removal from the waterbath, the device failed negative prep.On pressurisation of the device, a leak was noted on the balloon proximal pillow.Upon visual inspection, a short longitudinal tear was observed on the balloon proximal pillow.Lead in scratches were evident proximal and distal to the tear site.The balloon material was raised, jagged and uneven at the tear site.(b)(4).
|