(b)(6).(b)(4).Device evaluated by mfr: the catheter was received for analysis after decontamination.The returned device matches the upn and lot number provided by the customer.Visual inspection found a kink in the device approximately 13mm from the distal tip in the neutral position between r1 and r2.The ring 1 and ring 2 seals were found to be compromised; there is body fluid on the edge of the rings.Functional inspection found the steering knob and the tension control knob functioned properly on both lock and unlock positions.The catheter was placed on the curve template and the device passed both right and left curves tests.However, the device has a kink at the distal section at approximately 13mm from the tip.The distal section of the catheter was dissected.A small amount of body fluid was found in the interior of the sheath.The kevlar wrap was displaced and the center support was bent.One of the steering wires was detached, and there was evidence of solder on both the center support and the detached steering wire.The manufacturing batch record review confirmed that the device met all material, assembly and performance specifications.Complaint confirmed.The most probable root cause of the reported difficulty is a design constraint of the product.(b)(4).
|