MAUDE Adverse Event Report: AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AV-TEMECULA-CT NC TREK CORONARY DILATATION CATHETER Back to Search Results
Catalog Number 1012451-12
Device Problems Defective Device (2588); Device Misassembled During Manufacturing /Shipping (2912)
Patient Problem No Patient Involvement (2645)
Event Date 02/21/2017
Event Type  malfunction  
Manufacturer Narrative
(b)(4).During processing of this complaint, attempts were made to obtain complete event, patient and device information.The device was received.The investigation is not yet complete.A follow-up report will be submitted with all additional relevant information.
 
Event Description
Reportedly, a 3.5x12mm nc trek balloon dilatation catheter (bdc) was inspected prior to being used and it was noticed that the balloon and balloon markers didn't match.It is unknown if the markers were too close or the balloon was too long.The bdc was not used.The procedure was successfully completed with another 3.5x12mm nc trek bdc.There was no clinically significant delay in the procedure and no adverse patient effects.No additional information was provided.
 
Manufacturer Narrative
(b)(4).The device returned for analysis.The complaint investigation determined the reported difficulty was related to manufacturing issues associated with the protective sheath.On march 14, 2017, abbott vascular decided to initiate a voluntary field action for some sizes and lots of the nc trek family of dilatation catheters for difficulty to remove sheath which may lead to inflation or deflation issues.Abbott vascular performed a comprehensive investigation which included device analysis, manufacturing evaluation and trend analysis.The root cause identification was complicated by the fact that users were describing multiple symptoms when reporting the complaints.To date, the frequency of worldwide reported events for difficulties removing the protective balloon sheath, inflation and deflation has reached an actionable limit, thus abbott vascular communicated the voluntary field action to the fda [medwatch # 2024168-2017-02310].Corrective action has been implemented per site operating procedures.The product will continue to be trended.The abbott internal recall number is 2024168-3/14/2017-002-r.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NC TREK CORONARY DILATATION CATHETER
Type of Device
CORONARY DILATATION CATHETER
Manufacturer (Section D)
AV-TEMECULA-CT
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer (Section G)
EL COYOL, COSTA RICA REG# 3009031392
abbott vascular
26531 ynez road
temecula CA 92591 4628
Manufacturer Contact
connie speck
abbott vascular
26531 ynez road
temecula, CA 92591-4628
9519143996
MDR Report Key6412080
MDR Text Key70317755
Report Number2024168-2017-02231
Device Sequence Number1
Product Code LOX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K110134
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Health Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 03/31/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/16/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2019
Device Catalogue Number1012451-12
Device Lot Number60829G1
Was Device Available for Evaluation? Yes
Date Returned to Manufacturer03/01/2017
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/14/2017
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/01/2016
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-