Merge technical support assisted the customer with troubleshooting efforts.It was found that the coil cord and gender changer connector were faulty.The failed hardware was changed out with replacements the customer had on hand; the faulty hardware was not returned to merge healthcare for evaluation.The customer confirmed that the replacement hardware corrected the issue.Device labeling, hemo-5303 user manual - general equipment care, addresses the occurrence for such an issue with statements such as, "inspect overall physical condition of the system components, peripherals, and interconnecting cables.Perform any corrective actions required.".
|
Merge hemodynamics monitors, measures, and records physiological data from a human patient undergoing a cardiac catheterization procedure.The system comprises the patient data module and the merge hemodynamics hemo monitor pc.The two units are connected via a serial interface.All vital parameters and evaluations are registered and calculated in the patient data module.This data is then transmitted to the merge hemodynamics hemo monitor pc via the serial interface.All data can be shown and monitored on the merge hemodynamics hemo monitor pc.On (b)(6) 2017, a customer reported to merge healthcare that problems were experienced with the hemo monitor freezing during an emergent procedure.Subsequently, the hemo monitor was rebooted and resulted in a loss of patient monitoring.The delay was ~10-15 minutes while performing troubleshooting efforts; however, the procedure was completed successfully using an external defibrillator equipped with ekg and spo2.With merge hemo not capturing physiological data, there is a potential for incorrect treatment that could cause harm to the patient.However, the customer reported that the procedure was completed successfully using alternate means.(b)(4).
|