Model Number 70025S |
Device Problem
Break (1069)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/22/2017 |
Event Type
malfunction
|
Manufacturer Narrative
|
Cutter was returned in condition as described in the complaint.The end cap of the cutter became removed from the cutter during use.The end cap was returned in a separate pouch.No physical testing of performance was performed on the cutter, visual inspection only.Tube line set was cut off and needle had slight bend.Replacement cutter was also returned and a bend was noted in that cutter needle as well.Reviewed work order history for kit and no anomalies were noted.
|
|
Event Description
|
During procedure tip of the cutter broke off in the patient's eye.The doctor was able to retrieve the tip.Another cutter was used to complete the procedure.There was no patient injury reported.
|
|
Manufacturer Narrative
|
Correct spelling in manufactures address.Cutter was returned in condition as described in the complaint.The end cap of the cutter became removed from the cutter during use.The end cap was returned in a separate pouch.No physical testing of the performance was performed on the cutter, visual inspection only.The tube line set was cut off and the needle had a slight bend.A replacement cutter was also returned and a bend was noted in the cutter needle.Work order history was reviewed for the kit and no anomalies were noted.
|
|
Manufacturer Narrative
|
Vendor evaluation complete.Reviewed the cutter failure and noted that the weld failed on the end of the cutter causing the cap to come out of the end of the needle.It was noted that this was an isolated incident, the first in three years.Employee training was provided.
|
|
Search Alerts/Recalls
|