• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT IRELAND ARCHITECT AFP; ALPHA-FETOPROTEIN

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT IRELAND ARCHITECT AFP; ALPHA-FETOPROTEIN Back to Search Results
Catalog Number 03P36-26
Device Problem High Test Results (2457)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
An evaluation is in process.A follow-up report will be submitted when the evaluation is complete.
 
Event Description
The customer observed a falsely elevated afp result for one patient while using the architect afp assay.The customer provided the following data (unit of measure not provided): initial result: about 200, retest result: 2-3.There was no adverse impact to patient management reported.
 
Manufacturer Narrative
Further investigation of the customer issue included a review of the complaint text, trouble shooting of the analyzer, a search for similar complaints, and a review of labeling.Return material was not available.Review of the analyzer was completed and did not identify any issues.A trending review did not identify an adverse trend.Labeling was reviewed and found to be adequate.Based on the available information a deficiency of the architect afp reagent, list 03p36-26, was not identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCHITECT AFP
Type of Device
ALPHA-FETOPROTEIN
Manufacturer (Section D)
ABBOTT IRELAND
diagnostics division
finisklin business park
sligo NA
EI  NA
Manufacturer (Section G)
ABBOTT IRELAND
diagnostics division
finisklin business park
sligo NA
EI   NA
Manufacturer Contact
noemi romero-kondos, rn bsn
100 abbott park road
dept. 09b9, lccp1-3
abbott park, IL 60064-3537
224667-512
MDR Report Key6417501
MDR Text Key70412199
Report Number3008344661-2017-00022
Device Sequence Number1
Product Code LOK
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P120008
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Health Professional
Type of Report Initial,Followup
Report Date 05/15/2017
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/20/2017
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number03P36-26
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/19/2017
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Treatment
ARCHITECT I2000SR ANALYZER; LN 03M74-01, SN (B)(4)
-
-